Skip to main content
  • Research Article
  • Open access
  • Published:

The determination of Ochratoxin A based on the electrochemical aptasensor by carbon aerogels and methylene blue assisted signal amplification

Abstract

In this work, a novel aptamer-based electrochemical biosensor was developed for the determination of Ochratoxin A (OTA) by using carbon aerogels (CAs) and methylene blue (MB) as signal amplification strategy. CAs was used as carrier to load the abundant of complementary DNA (cDNA), which could enhance the hybridization between CAs-cDNA and aptamer immobilized on the electrode surface, thus provide more double-stranded DNA for MB intercalation. The current of MB on the CAs-cDNA/apt/AuE sensor was twice that on the cDNA/apt/AuE sensor, which indicated that the CAs with high surface area enabled a higher loading of the cDNA and absorbed more MB, thus realized the signal amplification strategy. The optimum experimental conditions including MB incubation time of 15 min, aptamer concentration of 4.0 μmol/L, hybridization time of 2.0 h, and OTA incubation time of 18 min were obtained. The change of peak current was linearly proportional to the OTA concentration in the range of 0.10–10 ng/mL with the actual detection limit of 1.0 × 10−4 ng/mL. The experimental results showed that the prepared CAs-cDNA/apt/AuE exhibited good specificity, acceptable reproducibility and repeatability. This sensor was applied to detect OTA in the spiked corn samples, and obtained an acceptable average recovery of 89%.

Background

Mycotoxins are toxic contaminants produced by the secondary metabolism of fungi, mainly saprophytic molds [1]. As one of the highly toxic mycotoxins, Ochratoxin A (OTA) secreted by Aspergillus and Penicillium has attracted much more attention because it contaminates broad range of agricultural products such as maize, wheat, rice, coffee, and peanut, then results in serious human and animal health problems including nephrotoxic, hepatotoxic, neurotoxic, teratogenic and immunotoxic activities [2]. So, it is increasingly necessary to develop a precise, rapid and low-cost method for OTA determination in various samples. Conventional instrumental analyses such as high performance liquid chromatography, liquid chromatography tandem mass spectrometry, and fluorescence are popular because of their high sensitivity, good accuracy and reproducibility [3,4,5,6]. However, they exist some drawbacks such as sophisticated equipment, high cost and requirement of technical skills [7]. The immunoassay methods based on antigen–antibody binding have the advantages of simple, rapid and easy to operation, and appear an useful tool for on-site detection of OTA [8,9,10,11]. However, the antibody preparation process is complex and time-consuming, high cost, and the antibody itself is unstable, immunogenic and false. So it can not be used as a final confirmation method, which hinders its wider application.

As a novel bio-recognition element, aptamers, single strand oligonucleotides, with the superiority including strong affinity, high stability, and easy modification of functional groups, have the potential designing highly sensitive, selective and structure switchable sensing assays [12,13,14]. Recently, aptamer-based electrochemical biosensors for OTA detection are prominent owing to their fast response, low cost, simple operation, easy to miniaturization of the instrument, and portability [15,16,17,18,19].

To realize the signal amplification and improve the sensitivity of the electrochemical aptasensors, nanomaterials have been chosen because their large specific surface area allows immobilizing more signal molecules on the electrode surface, and their well electronic conductivity makes the charge transfer to the electrodes easier [20,21,22,23]. Due to their favorable properties including great mesopore volume, high accessible surface area and good electrical conductivity, carbon aerogels (CAs) have attracted tremendous attention and have been extensively used as supports of precious metal for electrocatalytic reaction [24, 25], whereas have seldom been used for immobilization of biomolecules [26].

In this work, a novel aptamer-based electrochemical biosensor was developed for the determination of OTA by using CAs and methylene blue (MB) as signal amplification strategy. CAs was used as carrier to load the abundant of complementary DNA (cDNA), which could enhance the hybridization between CAs-cDNA and aptamer immobilized on the electrode surface, thus provide more double-stranded DNA for MB intercalation. As an electrochemical indicator, MB could intercalate both into single-stranded cDNA through the guanine bases and into double-stranded DNA, and produce a strong current signal. When OTA existed, the formation of aptamer-OTA complex changed the conformation of aptamer and prohibited the binding of cDNA-aptamer, which resulted in the release of MB from the electrode surface and produced a reduced current signal. The change of MB current signal could be used for OTA detection.

Methods

Materials and chemicals

1-Ethyl-3-(3-dimethylaminopropyl)-carbodiimide (EDC), N-hydroxysuccinimide (NHS), methylene blue (MB) were purchased by Macklin Biochemical Co., Ltd. (Shanghai, China). All oligonucleotides were synthesized by Sangong Biotech (Shanghai, China) Co., Ltd., and their base sequences were: complementary DNA (cDNA): 5′-NH2-GGA GGA GGA GGA GGA GGA GGA GGA GGA GGA GGA GGA GGA TGT CCG ATG CTC CCT TTA CGC CTC-3′; OTA aptamer (apt): 5′-HS-GAT CGG GTG TGGGTG GCG TAA AGG GAG CAT CGG ACA-3′. 50 mM, pH7.4 Tris–HCl was prepared by 0.20 M NaCl and 1.0 mM EDTA and adjusting the pH with 0.10 M HCl. All other chemicals were of analytical-reagent grade.

Apparatus

All the electrochemical experiments were performed on a CHI 660E Electrochemical Workstation (Shanghai Chenhua Instrument Corporation, China). A three-electrode system was comprised of Au electrode (AuE) as working electrode, platinum wire as auxiliary electrode, and Ag/AgCl as reference electrode. Scanning electron microscopy (SEM) was performed using a JEOL JSM7100F SEM facility (Jeol, Japan).

Preparation of the CAs-cDNA/apt/AuE sensor for OTA detection

The AuE was polished with 0.30 and 0.050 μm gamma alumina powder successively and then rinsed with ultrapure water and dried by nitrogen. The AuE was activated by scanning cyclic voltammogram (CV) with 0.50 M H2SO4.

CAs were synthesized by the sol–gel polymerization of resorcinol (R) and formaldehyde (F) in an aqueous solution according to the method described elsewhere [25]. cDNA (25 μL, 100 μM) was put into 500 μL of CAs suspension, then 250 μL of EDC and NHS was separately added into the solution, the mixture was incubated overnight at 37 °C. Next, the above solution was incubated with NaCl (50 μL, 2.0 M) for 24 h and centrifugated at 12,000 rpm to remove the unbound cDNA. The precipitate was repeatedly rinsed and redispersed in 5.0 mL Tris–HCl solution to obtain the CAs-cDNA.

5.0 μL of aptamer was immobilized on the surface of AuE to obtain apt/AuE via the S–Au bonds, and then the modified electrode was incubated with MCH to eliminate nonspecific binding and block the remaining active groups. Sequentially, 5.0 μL of CAs-cDNA was dropped on the apt/AuE surface and the hybridization reaction was proceeded at 37 °C to obtain the CAs-cDNA/apt/AuE sensor. The procedure of the aptasensor fabrication for OTA detection was illustrated in Scheme 1.

Scheme 1
scheme 1

Schematic illustration of the procedure of the CAs-cDNA/apt/AuE fabrication for OTA detection

Results and discussion

Characterization of the prepared CAs

The morphology of CAs was characterized by SEM, and the result was showed in Fig. 1a. It was observed that the prepared CAs had high surface area and three-dimensional interconnected porous structure. The sizes of particles have been found to range from 50 nm to 100 nm and uniformly distributed. The pore sizes between particles were in the range of 20–150 nm. Figure 1b showed the nitrogen adsorption–desorption isotherms of CAs. It displayed an obvious hysteresis loop, indicating the presence of mesopores in CAs. The surface area and pore volume of CAs were measured on Autosorb IQ (Quantachrome) using the Brunauer–Emmett–Teller (BET) method. The BET surface area and pore volume of CAs were 695 m2 g−1 and 0.90 cm3 g−1, respectively. These high surface area and porous structure made CAs expose more active sites, accelerate the transfer rate and improve the electrochemical performance.

Fig. 1
figure 1

a SEM image and b N2 adsorption–desorption isotherms of CAs

Electrochemical characterization of the CAs-cDNA/apt/AuE sensor

Electrochemical impedance spectroscopy was used to characterize the different electrodes, and the results were shown in Fig. 2. For the bare AuE (a), the charge transfer resistance (Rct) was 80 Ω, indicating that the bare AuE had good conductivity. When the aptamer was modified on the AuE surface to obtain the apt/AuE (b), the Rct was 831 Ω, which increased obviously as compared to that of the bare AuE. This was due to that the repulsion between the negatively charged backbones of DNA strands and [Fe(CN)6]3−/4− hindered the interfacial electron transfer. For the CAs-cDNA/apt/AuE (c), after the CAs-cDNA hybridized with the aptamer, the negative charge density of the electrode surface further increased, so the Rct significantly increased to 1491 Ω. These results were also proved that the aptamer and the CAs-cDNA had been successfully immobilized on the electrode surface. On the basis of the charge transfer kinetics of the [Fe(CN)6]3−/4−, the Faradaic impedance spectra were modeled using the Randles equivalent circuit (inset of Fig. 2). The fitting parameters involved the resistance of the solution (Rs), the electron-transfer resistance (Rct), Warburg impedance (Zw) attributed to the contribution of diffusion, and the constant phase element (Q).

Fig. 2
figure 2

The EIS of 10.00 mM [Fe(CN)6]3−/4− on the different electrodes. (a) The bare AuE, (b) the apt/AuE, (c) the CAs-cDNA/apt/AuE. Inset: the equivalent circuit

Electrochemical behavior of MB on the different sensors

Using MB as the electrochemical probe, the cDNA/apt/AuE (a) and the CAs-cDNA/apt/AuE (b) were characterized by the differential pulse voltammetry (DPV). As shown in Fig. 3, the peak current of MB was 2.7 μA on the cDNA/apt/AuE and 5.4 μA on the CAs-cDNA/apt/AuE. The peak current on the CAs-cDNA/apt/AuE was twice that on the cDNA/apt/AuE, which was ascribed to that the high surface area and porous structure of CAs made it load more cDNA and absorb more MB, thus realize the signal amplification.

Fig. 3
figure 3

DPV responses of different electrode in Tris–HCl buffer after incubated in 0.02 μM MB. (a) The cDNA/apt/AuE, (b) the CAs-cDNA/apt/AuE

The detection mechanism of OTA based on the CAs-cDNA/apt/AuE sensor

Figure 4 showed the signal change of 0.02 μM MB before and after incubation with 10 ng/mL OTA. The peak current of MB on the CAs-cDNA/apt/AuE (a) was 5.4 μA and obviously decreased to 3.8 μA on the OTA/CAs-cDNA/apt/AuE (b). The ΔI was 1.6 μA before and after incubation with OTA, which can be applied for OTA detection. This signal change can be explained as follows: In the absence of OTA, MB can intercalate both into single-stranded cDNA through the guanine bases and into double-stranded DNA, thus produce a strong current signal. In the presence of 10 ng/mL OTA, the binding of OTA and aptamer is considerably greater than that of cDNA and aptamer, which results in the release of MB from the electrode surface and produces a reduced current signal.

Fig. 4
figure 4

The DPV of the CAs-cDNA/apt/AuE in Tris–HCl buffer after incubated in 0.02 μM MB. (a) 0 ng/mL OTA, (b) 10 ng/mL OTA

The optimization of the important factors

The effect of incubation time of MB on the CAs-cDNA/apt/AuE sensor was studied. As shown in Fig. 5a, the peak current obviously increased with the increase of incubation time from 5 min to 15 min. The uptrend was slowly when the incubation time exceeded 15 min. So 15 min of incubation time was used in the following experimental study. Figure 5b showed the optimized results of the aptamer concentration. It can be seen that the peak current of MB obviously increased when the aptamer concentration increased from 2.0 to 4.0 μmol/L, and reached a maximum of 5.8 μA at 4 μmol/L. Beyond the aptamer concentration of 4.0 μmol/L, the current signal decreased slowly, this is because that the excess aptamer immobilized on the electrode surface could hinder the interfacial electron transfer. Thus, 4.0 μmol/L was used. Figure 5c showed the optimized results of the hybridization time between cDNA and aptamer. The current signal increased dramatically in the range of 1.0–2.0 h, and reached a platform when the hybridization time exceeded 2.0 h, indicating the amount of DNA on the electrode surface was saturated. The dependence of the ΔI on the incubation time of OTA was also optimized. As shown in Fig. 5d, it can be seen that the ΔI changed obviously when the incubation time was before 18 min and then changed slowly after 18 min. In order to reduce the detection time for the prepared sensor, 18 min of the incubation time of OTA was used in the further experiments. Figure 5e showed the optimized results of the concentration of CAs-cDNA. The current signal increased dramatically when the concentration of CAs-cDNA increased from 2.0 to 5.0 μmol/L, and reached a maximum of 5.4 μA at 5.0 μmol/L. The current signal decreased when the concentration of CAs-cDNA exceeded 5.0 μmol/L. This was because that the excess CAs-cDNA immobilized on the electrode surface could hinder the interfacial electron transfer.

Fig. 5
figure 5

The effect of a the incubation time of MB, b the aptamer concentration, c the hybridization time between cDNA and aptamer, d the incubation time of OTA, and e the concentration of CAs-cDNA on the CAs-cDNA/apt/AuE sensor

Analytical performance of the CAs-cDNA/apt/AuE sensor

Using the optimized parameters, different concentrations of OTA were detected on the CAs-cDNA/apt/AuE sensor. As shown in Fig. 6a, the increase in ΔI was observed upon increasing OTA concentration in the range 1.0 × 10−4–500 ng/mL. As shown in Fig. 6b, the ΔI was linearly proportional to OTA concentration in the range of 0.10–10 ng/mL, and the linear regression equation was ΔI = − 0.04x − 1.37 (R2 = 0.995). The detection limit of the proposed aptasensor was 1.0 × 10−4 ng/mL. Compared with the reported literatures, the CAs-cDNA/apt/AuE sensor was superior to other aptasensors, and the results were shown in Table 1.

Fig. 6
figure 6

a The dependence of ΔI on increasing OTA concentrations. b The linear relationship between ΔI and OTA concentrations

Table 1 Comparison with other reported aptasensors for OTA detection

The specificity of the CAs-cDNA/apt/AuE sensor

In order to investigate the specificity of the as-prepared aptasensor, control experiments were performed using AFB1 and ZEA. As shown in Fig. 7, the peak current response from incubation with AFB1 and ZEA did not produce obvious variations compared with that from the blank solution, while induced the great decrease after incubation with OTA. These results demonstrated that the proposed aptasensor had good specificity towards OTA detection.

Fig. 7
figure 7

The specificity of the developed aptasensor towards OTA

Reproducibility and repeatability of the CAs-cDNA/apt/AuE sensor

The reproducibility of the developed CAs-cDNA/apt/AuE sensor was evaluated with inter-assay precision. The five CAs-cDNA/apt/AuE sensors were tested for DPV with same OTA concentration under the same experimental conditions. A relative standard deviation (RSD) of 6.7% was calculated, indicating a good reproducibility of the developed aptasensor. The intra-assay precision of the CAs-cDNA/apt/AuE sensor was evaluated by five repetitive measurements with one electrode and RSD of 7.3% was obtained, indicating that the prepared aptasensor had acceptable repeatability. The prepared CAs-cDNA/apt/AuE was stored at 4 °C when not in use. After a 20-day storage period, the sensor retained 93% of its initial current response, providing the acceptable stability.

The application of the aptasensor to corn sample

To investigate the actual performance of the developed aptasensor, the OTA concentration in spiked corn sample was examined. As shown in Table 2, the recovery was in range of 86–93% and the average recovery was 89%. This implied that the as-prepared aptasensor had a promising feature for the practical use in corn sample.

Table 2 The detection of OTA in the spiked corn sample

Conclusions

In this work, a novel CAs-cDNA/apt/AuE sensor was developed to detect OTA using CAs and MB assisted signal amplification. The CAs could load the abundant cDNA and absorb more MB, so the peak current of MB on the CAs-cDNA/apt/AuE sensor was higher than that on the cDNA/apt/AuE sensor. Under the optimized experimental conditions, the developed aptasensor could detect OTA at the level of 1.0 × 10−4 ng/mL, and exhibited good specificity against ZEA and AFB1. This sensor was also applied to detect OTA in the spiked corn samples, and an acceptable average recovery of 89% was obtained. By changing the aptamers for different target molecules, this strategy has potential prospect for detecting other targets in the convenient field monitoring.

References

  1. Vidal JC, Bonel L, Ezquerra A, Hernández S, Bertolín JR, Cubel C, Castillo JR (2013) Electrochemical affinity biosensors for detection of mycotoxins: a review. Biosens Bioelectron 49:146–158

    Article  CAS  Google Scholar 

  2. Rhouat A, Yang C, Hayat A, Marty JL (2013) Aptamers: a promising tool for Ochratoxin A detection in food analysis. Toxins 5:1988–2088

    Article  Google Scholar 

  3. Wu SJ, Duan N, Ma XY, Xia Y, Wang HX, Wang ZP, Zhang Q (2012) Multiplexed fluorescence resonance energy transfer aptasensor between upconversion nanoparticles and graphene oxide for the simultaneous determination of mycotoxins. Anal Chem 84:6263–6270

    Article  CAS  Google Scholar 

  4. Ahn S, Lee S, Lee J, Kim B (2016) Accurate determination of Ochratoxin A in Korean fermented soybean paste by isotope dilution-liquid chromatography tandem mass spectrometry. Food Chem 190:368–373

    Article  CAS  Google Scholar 

  5. Paschoal FN, Silva DDA, Souz RVSD, Oliveira MSD, Pereira DAA (2017) A rapid single-extraction method for the simultaneous determination of aflatoxins B1, B2, G1, G2, fumonisin B1, and zearalenone in corn meal by ultra performance liquid chromatography tandem mass spectrometry. Food Anal Methods 10:1631–1644

    Article  Google Scholar 

  6. Savastano ML, Losito I, Pati S (2016) Rapid and automatable determination of Ochratoxin A in wine based on microextraction by packed sorbent followed by HPLC-FLD. Food Control 68:391–398

    Article  CAS  Google Scholar 

  7. Tai HH (2015) Recent advances for the detection of Ochratoxin A. Toxins 7:5276–5300

    Article  Google Scholar 

  8. Wu SJ, Duan N, Zhu CQ, Ma XY, Wang M, Wang ZP (2011) Magnetic nanobead-based immunoassay for the simultaneous detection of aflatoxin B1 and Ochratoxin A using upconversion nanoparticles as multicolor labels. Biosens Bioelectron 30:35–42

    Article  CAS  Google Scholar 

  9. Liu X, Xu Y, Wan DB, Xiong YH, He ZY, Wang XX, Gee SJ, Ryu D, Hammock BD (2015) Development of a nanobody—alkaline phosphatase fusion protein and its application in a highly sensitive direct competitive fluorescence enzyme immunoassay for detection of Ochratoxin A in cereal. Anal Chem 87:1387–1394

    Article  CAS  Google Scholar 

  10. Liu X, Tang ZW, Duan ZH, He ZY, Shu M, Wang XX, Gee SJ, Hammock BD, Xu Y (2017) Nanobody-based enzyme immunoassay for Ochratoxin A in cereal with high resistance to matrix interference. Talanta 164:154–158

    Article  CAS  Google Scholar 

  11. Sun ZC, Duan ZH, Liu X, Deng X, Tang ZW (2017) Development of a nanobody-based competitive dot ELISA for visual screening of Ochratoxin A in cereals. Food Anal Methods 10:3558–3564

    Article  Google Scholar 

  12. Hayat A, Yang C, Rhouati A, Marty JL (2013) Recent advances and achievements in nanomaterial-based, and structure switchable aptasensing platforms for Ochratoxin A detection. Sensors 13:15187–15208

    Article  CAS  Google Scholar 

  13. Wang CQ, Qian J, Wang K, Wang K, Liu Q, Dong XY, Wang CK, Huang XY (2015) Magnetic-fluorescent-targeting multifunctional aptasensorfor highly sensitive and one-step rapid detection of Ochratoxin A. Biosens Bioelectron 68:783–790

    Article  CAS  Google Scholar 

  14. Duan N, Wu SJ, Dai SL, Gu HJ, Hao LL, Ye H, Wang ZP (2016) Advances in aptasensors for the detection of food contaminants. Analyst 141:3942–3961

    Article  CAS  Google Scholar 

  15. Sun AL, Zhang YF, Sun GP, Wang XN, Tang D (2015) Homogeneous electrochemical detection of Ochratoxin A in foodstuff using aptamer–graphene oxide nanosheets and DNase I-based target recycling reaction. Biosens Bioelectron 89:659–665

    Article  Google Scholar 

  16. Catanante G, Mishra RK, Hayat A, Marty JL (2016) Sensitive analytical performance of folding based biosensor using methylene blue tagged aptamers. Talanta 153:138–144

    Article  CAS  Google Scholar 

  17. Mejri-Omrani N, Miodek A, Zribi B, Marrakchi M, Hamdi M, Marty JL, Korri-Youssoufi H (2016) Direct detection of OTA by impedimetric aptasensor based on modified polypyrrole-dendrimers. Anal Chim Acta 920:37–46

    Article  CAS  Google Scholar 

  18. Mishra RK, Hayat A, Gaelle C, Istamboulie G, Marty JL (2016) Sensitive quantitation of Ochratoxin A in cocoa beans using differential pulse voltammetry based aptasensor. Food Chem 192:799–804

    Article  CAS  Google Scholar 

  19. Liu C, Guo YJ, Luo F, Rao PF, Fu CL, Wang SY (2017) Homogeneous electrochemical method for Ochratoxin A determination based on target triggered aptamer hairpin switch and exonuclease III-assisted recycling amplification. Food Anal Methods 10:1982–1990

    Article  Google Scholar 

  20. Wu L, Xiong E, Zhang X, Zhang XH, Chen JH (2014) Nanomaterials as signal amplification elements in DNA-based electrochemical sensing. Nano Today 9:197–211

    Article  CAS  Google Scholar 

  21. Rivas L, Mayorgamartinez CC, Quesada-González D, Zamora-Gálvez A, Escosura-Muñiz ADL, Merkoçi A (2015) Label-free impedimetric aptasensor for Ochratoxin-A detection using iridium oxide nanoparticles. Anal Chem 87:5167–5172

    Article  CAS  Google Scholar 

  22. Wang CQ, Qian J, An K, Huang XY, Zhao LF, Liu Q, Hao N, Wang K (2016) Magneto-controlled aptasensor for simultaneous electrochemical detection of dual mycotoxins in maize using metal sulfide quantum dots coated silica as labels. Biosens Bioelectron 89:802–809

    Article  Google Scholar 

  23. Wei M, Feng S (2017) A signal-off aptasensor for the determination of Ochratoxin A by differential pulse voltammetry at a modified Au electrode using methylene blue as an electrochemical probe. Anal Methods 9:5449–5454

    Article  CAS  Google Scholar 

  24. Rooke J, Passos CDM, Chatenet M, Sescousse R, Budtova T, Berthon-Fabry S, Mosdale R, Maillard F (2011) Synthesis and properties of platinum nanocatalyst supported on cellulose-based carbon aerogel for applications in PEMFCs. J Electrochem Soc 158:B779–B789

    Article  CAS  Google Scholar 

  25. Calderón JC, Mahata N, Pereira MFR, Figueiredo JL, Fernandes VR, Rangel CM, Calvillo L, Lázaro MJ, Pastor E (2012) Pt–Ru catalysts supported on carbon xerogels for PEM fuel cells. Int J Hydrogen Energy 37:7200–7211

    Article  Google Scholar 

  26. Liu Y, Wei M (2014) Development of acetylcholinesterase biosensor based on platinume carbon aerogels composite for determination of organophosphorus pesticides. Food Control 36:49–54

    Article  Google Scholar 

  27. Tong P, Zhang L, Xu JJ, Chen HY (2011) Simply amplified electrochemical aptasensor of Ochratoxin A based on exonuclease-catalyzed target recycling. Biosens Bioelectron 29:91–101

    Article  Google Scholar 

  28. Jiang L, Qian J, Yang XW, Yan YT, Liu Q, Wang K, Wang K (2014) Amplified impedimetric aptasensor based on gold nanoparticles covalently bound graphene sheet for the picomolar detection of Ochratoxin A. Anal Chim Acta 806:128–135

    Article  CAS  Google Scholar 

  29. Qian J, Jiang L, Yang XW, Yan YT, Mao HP, Wang K (2014) Highly sensitive impedimetric aptasensor based on covalent binding of gold nanoparticles on reduced graphene oxide with good dispersity and high density. Analyst 139:5587–5593

    Article  CAS  Google Scholar 

  30. Yang ML, Jiang BY, Xie JQ, Xiang Y, Yuan R, Chai YQ (2014) Electrochemiluminescence recovery-based aptasensor for sensitive Ochratoxin A detection via exonuclease-catalyzed target recycling amplification. Talanta 125:45–50

    Article  CAS  Google Scholar 

Download references

Authors’ contributions

MW planed and supervised the whole work, and revised the manuscript. WYZ carried out the experiments and drafted the manuscript. Both authors read and approved the final manuscript.

Acknowledgements

This study was funded by the Natural Science Foundation of Henan Province (182300410188), the Fundamental Research Funds for the Henan Provincial Colleges and Universities in Henan University of Technology (2016RCJH04).

Competing interests

The authors declare that they have no competing interests.

Ethics approval and consent to participate

Not applicable.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Min Wei.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Wei, M., Zhang, W. The determination of Ochratoxin A based on the electrochemical aptasensor by carbon aerogels and methylene blue assisted signal amplification. Chemistry Central Journal 12, 45 (2018). https://doi.org/10.1186/s13065-018-0415-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13065-018-0415-4