An industry consensus study on an HPLC fluorescence method for the determination of (±)-catechin and (±)-epicatechin in cocoa and chocolate products

Abstract

Background

This manuscript describes the results of an HPLC study for the determination of the flavan-3-ol monomers, (±)-catechin and (±)-epicatechin, in cocoa and plain dark and milk chocolate products. The study was performed under the auspices of the National Confectioners Association (NCA) and involved the analysis of a series of samples by laboratories of five member companies using a common method.

Methodology

The method reported in this paper uses reversed phase HPLC with fluorescence detection to analyze (±)-epicatechin and (±)-catechin extracted with an acidic solvent from defatted cocoa and chocolate. In addition to a variety of cocoa and chocolate products, the sample set included a blind duplicate used to assess method reproducibility. All data were subjected to statistical analysis with outliers eliminated from the data set.

Results

The percent coefficient of variation (%CV) of the sample set ranged from approximately 7 to 15%.

Conclusions

Further experimental details are described in the body of the manuscript and the results indicate the method is suitable for the determination of (±)-catechin and (±)-epicatechin in cocoa and chocolate products and represents the first collaborative study of this HPLC method for these compounds in these matrices.

Background

The use of cocoa has been documented for almost 4,000 years. The first population thought to consume the material was the Mesoamericans [1, 2]. In the past decade there has been increasing interest and numerous publications on the putative health effects associated with the moderate consumption of cocoa and chocolate products [35]. In a parallel fashion, several groups initiated studies into the potential agents responsible for cardiovascular effects with the flavanols, (±)-catechin and (±)-epicatechin being candidate compounds [6].

The growing interest in these compounds resulted in a plethora of methods for quantification in various foodstuffs including tea, wine, grapes and chocolate. While other analytical methods have been used, HPLC was the predominant method developed [713]. A thorough literature search using Google and Pubmed resulted in thousands of citations on polyphenol analysis and almost 900 citations on flavanol analysis by HPLC in chocolate, indicating the recent explosive growth in methods for these analytes. Considering the increased interest in the cocoa flavanols' potential cardiovascular effects, a standard quantification method would be pertinent for accurate determination of dose-response effects in clinical trials. There is both an ISO and an Institute for Nutraceutical Advancement (INA) method for flavanols in tea, but not yet a standard method for flavanol quantification in chocolate and cocoa [14, 15].

With this as background, the National Confectioners Association (NCA) convened an analytical chemistry working group to develop a consensus HPLC method. This group conducted a collaborative study using samples provided by NCA to establish a method to quantify (±)-catechin and (±)-epicatechin in cocoa and chocolate and make it available to the industry.

Experimental

Scope and Applicability

This method is applicable for the analysis of (±)-epicatechin and (±)-catechin in cocoa powder, chocolate liquor and formulated chocolate products. This ring trial only included pure chocolate; any products containing inclusions (such as fruit or nuts) may not be appropriate for this method due to potential interference.

A. Principle

This method determines the (±)-catechin and (±)-epicatechin content of cocoa and chocolate products. Fat is removed from the sample in order to prevent potential interference and protect the column by using multiple hexane extractions. Defatted samples are then dried for subsequent extraction of analytes. Defatted, dried samples are extracted, with sonication, at 40°C for 15 minutes using an acetone: water: acetic acid (70: 29.5: 0.5) solvent mixture. Extracted samples are then centrifuged to remove insoluble materials and brought up to a defined volume. The extracts are filtered into HPLC vials for chromatographic analysis. (±)-Catechin and (±)-epicatechin are separated by a reverse phase mechanism on a C18 column with an acidic acetonitrile-water mobile phase gradient. Analytes are detected and quantified by their fluorescence, with excitation at 280 nm and emission at 315 nm.

B. Apparatus

  • (a) HPLC system: With solvent degasser, binary gradient pumping, gradient mixer, injector capable of 10 µL injection (either autosampler or manual), column oven, fluorescence detector and data analysis system

  • (b) Chromatography column: Reversed phase HPLC column octadecylsilane (ODS; C18) derivatized silica reversed phase HPLC column, pore size from 100 - 125 A, are recommended. Recommended column: Phenomenex Luna, 5 µm, C18(2), 100A, 250 × 3.0 mm (alternate columns may be used if they provide acceptable resolution)

  • (c) Analytical balance: Readability 0.1 mg or lower

  • (d) Pipettes: Capable of accurately delivering 20-1000 µL; 1-5 mL

  • (e) Vials: 2 mL, amber glass, screw cap, for storing Stock Standard solutions and for holding filtered HPLC sample prior to injection

  • (f) Test Tubes: Screw capped, with caps, capable of holding at least 10 mL

  • (g) Volumetric flasks: 10 mL, 20 mL, 50 mL and 100 mL, Class A, glass

  • (h) Centrifuge tubes: Plastic, for single use, 50 mL, screw cap (air tight)

  • (i) Vortex Mixer

  • (j) Ultrasonic Bath

  • (k) Flame-Proof Centrifuge: For centrifuging 50 mL tubes at 2500 × g

  • (l) Syringe filters: For filtering HPLC samples, 0.45 µm PVDF, PTFE or hydrophilic polypropylene, 13 or 25 mm diameter (Nylon filters are not recommended due to potential adsorption of metabolites)

  • (m) Syringe: All plastic, 1 mL to 5 mL as appropriate

  • (n) Glass beads: (approx. diameter 5 mm)

C. Reagents

  • (a) Water: High purity deionized water, filtered through a 0.45 µm or smaller pore filter

  • (b) Acetonitrile: HPLC grade

  • (c) Hexane: HPLC grade

  • (d) Acetone: HPLC grade

  • (e) Acetic Acid: Glacial

  • (f) Extraction Solvent: Mix 700 mL Acetone, 295 mL Water and 5 mL Acetic Acid

  • (g) Mobile Phase A: 0.2% Acetic Acid in Water. Add 2 mL Acetic Acid to 1 L Water.

  • (h) Mobile Phase B: 0.2% Acetic Acid in Acetonitrile. Add 2 mL Acetic Acid to 1 L Acetonitrile.

  • (i) Standards: (±)-Catechin hydrate, purity = 98%, Sigma-Aldrich C1251-5G or equivalent; (±)-epicatechin, purity = 98%, Sigma-Aldrich E4018-1G or equivalent. Certificate of analysis from supplier is required for purity correction of each new lot number.

D. Standards and Reagent Blank Preparation

  • (a) (i.) Stock standard solution A (approx 1000 µg/mL): Into a 50 mL volumetric flask, accurately weigh approximately 50 mg (±)-catechin hydrate and 50 mg (±)-epicatechin and record the weights. Add extraction solvent and mix or sonicate to dissolve. Bring to 50.00 mL with extraction solvent and mix. Label as Stock standard A.

  • (ii.) Stock standard solution B (approx 100 µg/mL): Pipette, using a Class A volumetric pipette, 5 mL of Stock standard A into a 50 mL volumetric flask and dilute to volume with extraction solvent. Label as Stock standard B.

  • (b) Obtain loss on drying (important for (±)-catechin hydrate): Crystal water is not stoichiometrically distributed; for (±)-epicatechin loss on drying normally equals 0) and HPLC purity of analyte from supplier's certificate of analysis for each new lot number to calculate purity: Purity (%) = [100 (%) - loss on drying (%)] * HPLC purity (%)/100 (%)

  • Calculate exact concentrations of each component in stock standard solution as shown below:

Stock standard A use concentration as is.

Stock standard B concentrations will require multiplication by an additional 1/10 factor.

  • (c) Fluorescence Detector Sensitivity Assessment: Stock Standard Selection (Stock Standard Solutions A, B): When running this method for the first time inject the appropriate injection volume, 10 µl, of Stock Standard A and B onto the HPLC system under the conditions provided in Section G. Examine the detector response for the two concentrations provided. Choose the most concentrated stock standard solution that does not saturate the detector as the Stock Standard with which to proceed. Discard the other stock standard solution. If proceeding with stock standard B, for future analysis note that the stock standard preparation procedure can be modified by preparing a stock standard solution of 0.1 mg/mL to save one dilution step.

  • (d) Reagent Blank: Use extraction solvent for the blank.

  • (e) Working standards: Prepare Working Standard Solutions from the chosen stock standard. Add indicated amounts of appropriate working standard (100 or 1000 µg/mL) to a 10 mL volumetric flask; bring to volume with the extraction solvent. Transfers should always be made with Class A volumetric pipettes. Alternatively test tubes can be used and the remaining extraction solvent for dilution to 10 mL can be added with Class A volumetric pipettes. See an example of the working standard dilution scheme in Table 1 below.

    Table 1 Example of Working Standard Dilution Scheme(s)
  • (f) Calculate the exact concentration of each component of the working standards as follows:

E. Lipid Removal from Cocoa and Chocolate Samples

  • (a) Accurately weigh approximately 2 grams of each finely divided/grated milk chocolate sample or 1 gram for cocoa powders/baking chocolate/dark chocolate samples into a labeled, tared 50 mL disposable centrifuge tube. Record the weight of the sample W SAMPLE.

  • (b) Add approx. 40 mL hexane (dispenser) and cap tightly.

  • (c) Mix until the sample is completely dispersed (check visually).

  • (d) Centrifuge for 5 minutes at 2500 × g.

  • (e) Carefully decant and dispose of the hexane phase immediately.

  • (f) Repeat defatting steps (b) to (e) one additional time.

  • (g) Remove the cap and allow the residual solvent to evaporate in an appropriate fume hood until remaining hexane has evaporated (e.g. over night). Alternately, a stream of nitrogen may be used to accelerate the drying process.

F. Preparation of Test Solutions

  • Continue with whole sample remaining in the centrifugation tube.

  • (a) Add 2 glass beads to the centrifuge tube containing the dried, defatted sample.

  • (b) Add 9 mL of extraction solvent (dispenser) and vigorously shake the sample to break centrifugation pellet. Sample does not need to be completely suspended yet. Shake headlong, if necessary gently tap several times.

  • (c) Place in an ultrasonic bath at 40°C for 15 minutes in total. After 5-10 minutes of sonication, remove sample from bath and handshake again until sample is completely suspended (check visually). Alternately, vortex sample.

  • (d) Remove the sample from the ultrasonic bath, centrifuge at 2500 × g for 5 minutes.

  • (e) Carefully and slowly decant the liquid portion into a 20 mL Class A volumetric flask (wide neck, if possible).

  • (f) Repeat the extraction steps (b) to (d) one additional time. Decant the liquid from the second extraction into the same 20 mL volumetric flask.

  • (g) Bring to volume with extraction solvent.

  • (h) Assemble a Syringe and Syringe Filter. Filter approximately 1 mL of sample into a HPLC Vial.

  • (i) Analyze by HPLC as described in Section G.

G. Chromatography

  • (a) Injection volume: 10 µL

  • (b) Flow rate: 0.65 mL/min for 3 mm i.d. column; Alter flow rate to maintain linear flow for other column dimensions.

  • (c) Detection: Fluorescence with excitation at 280 nm and emission at 315 nm

  • (d) Column temperature: 40°C

  • (e) Gradient Elution: See Table 2for example gradient conditions. HPLC columns differ in their selectivity for these compounds. Gradient conditions should be altered as needed to achieve resolution of (±)-catechin and (±)-epicatechin from interfering peaks.

    Table 2 HPLC Gradient Example
  • (f) Concentration of analytes in sample extract: Check if concentration of analytes in sample extract lie within their calibration ranges. If necessary, dilute and re-run extraction solution.

  • (g) Check sample: Check sample by re-runing mid point calibration curve in middle and at end of sample sequence. Calculate the mean, standard deviation and coefficient of variation (%CV) of the peak areas.

H. Calculations

Integrate peak area for quantitation. If peak areas of analytes in sample extracts are above calibration curve, dilute sample extract solution with extraction solvent accordingly. If peak areas of analytes in sample extracts are below calibration curve, repeat sections E and F and increase sample weight accordingly.

Construct standard curves, plotting calibration standard concentration of each standard against the area of the standard peak, using linear regression. Calculate the analytes (±)-catechin and (±)-epicatechin in the original sample as follows:

Analyte in sample [µg/g] = assay concentration of analyte [µg/mL] * × [mL]/W sample [g]

WSample [g] = initial sample weight from section E (a)

× [mL] = 20 mL (volume of extraction solution in volumetric flask; section F)

I. HPLC System and Column Performance Criteria Qualification

An HPLC column which fully resolves the analytes of interest may be used for the method. Gradient slope, flow rates and injection volumes may be altered as appropriate to accommodate columns of differing dimensions.

J. System Suitability

System suitability is a required procedure to ensure the HPLC system is working correctly. The following suitability tests are recommended to ensure correct system operation prior to initial use:

Repeatability and carry-over: Before running any test solutions, demonstrate the repeatability and lack of carryover of the HPLC system as follows:

  • (a) System Artifacts: As the first two injections of the day, analyze the blank standard twice in succession. Inspect the two chromatograms for artifact peaks from the HPLC system. Artifacts in the first chromatogram, absent in the second, indicate a buildup of impurities on the system. Artifacts present in both runs indicate impurities expected in every run. If the first chromatogram shows artifact peaks but the second chromatogram does not, inject a blank solution as the first sample in every analytical set. The presence of artifact peaks indicates impurities in the HPLC solvents, the needle wash system, or carryover in the injection system. These problems, if present, should be corrected.

  • (b) Carryover: Inject Standard 5 and then the blank. Carefully examine the blank injection for carryover peaks. Calculate the carryover of any peaks seen in the blanks as a percentage of the concentration found in standard 5. Carryover of standard 5 to the blank injection should be less than 0.1%.

  • (c) Linearity of the standard curve: Analyze each of the 5 standards, and construct a standard curve. The R2 of each standard curve should be greater than 0.9990. If this linearity is not achieved, prepare fresh standards.

  • (d) System precision: Analyze five replicate analyses of standard number 3. Calculate the concentration of each analyte and calculate the mean, standard deviation and percent coefficient of variation (%CV) of the results. The %CV for all peaks should be ideally less than 2%.

K. Samples

Samples for analysis were prepared by the NCA Study Director and submitted as blind samples to five participating laboratories. Samples consisted of cocoa, dark chocolate, milk chocolate and NIST SRM [2384] Baking Chocolate having certified values for (±)-catechin and (±)-epicatechin. One dark chocolate samples was used as a blind duplicate to assess method repeatability.

L. Quantification

Quantification was performed using the external calibration method as described in the method with all laboratories reporting regression coefficients in excess of 0.99 with the labs equally divided whether calibration was forced through zero.

Results and Discussion

All results were submitted to the Study Director using the form that was provided with the samples and with all data statistically evaluated. Samples were run in duplicate or triplicate. Furthermore each data set was evaluated using the Q-test to test for outliers with some data being eliminated. The results can be seen in Table 3.

Table 3 Flavan-3-ol Quantification

Figure 1 is a sample chromatogram provided from one laboratory of a dark chocolate extract. Instrumental conditions are as described in the methods section.

Figure 1
figure 1

Chromatogram of Dark Chocolate Extract Analyzed Under Method Parameters. Peak at 11.964 is (±)-catechin while peak at 14.422 is (±)-epicatechin.

Each laboratory also provided information about LOD (Limit of Detection) which was in pure solvent in the 40-50 ng/mL. The LOQ (Limit of Quantitation) ranged from 1- 2 μg/g Repeat injections of standards were also accomplished with %CVs reported in the 1-4% range.

The NIST (National Institute of Standard Technology) reference values for (±)-catechin and (±)-epicatechin in SRM (Standard Reference Material) 2384 are 245 +/- 51 (μg/g) and 1220 +/- 22 (μg/g) respectively with the data from this study indicating value of 254 +/- 23.5 (μg/kg) and 1137 +/- 68 (μg/kg) which are within the acceptable range of determinations established by NIST. While recovery studies have become a default method to assess method accuracy according to Swartz and Krull, the analysis of an established SRM is by itself a generally accepted method of validation [17]. Additionally, guidance from AOAC on methods validation indicates that spiking is not a desirable method to assess method accuracy as spiking solutions tend to be easily extractable hence the choice of the NIST standard to evaluate the method.

The sample labeled Milk Chocolate 1 is an example for a product containing very low amounts of the target analytes. With the analyte concentration in the sample extract at their lower limit of quantification and the chromatographic performance negatively affected by co-extracted matrix compounds the applied method operates at its limit. Hence the sample was not included in the statistical evaluation of the method. That being said, no issues related to complexation of polyphenols with milk reported by some researchers were seen [18].

The %CV ranged from 7-15% in this study. Laboratories used a column that satisfied the requirements of U.S. Pharmacopeia, previously described in methodology section [19]. The method was reviewed and compared with recommendations of Swartz and Krull. Finally, while the data in Table 3 may seem excessive to the casual observer, it is well within the parameters established by AOAC for another complex analyte [20].

The literature reports on the use of numerous solvents for the extraction of flavan-3-ols including mixtures of methanol, acetone, water and acid therefore the solvent combination used is within established parameters [21, 22]. Furthermore, a variety of HPLC detector types have been used including UV, Diode Array, Mass Spec and fluorescence [2327]. The choice of fluorescence detection is within established analytical parameters for this determination as it offers selectivity and sensitivity for these compounds with the identity of the peaks being established by the use of authentic standards.

Conclusion

The data from these studies indicate the proposed chocolate and cocoa method is suitable as an HPLC method for the determination of flavanol monomers, (±)-catechin and (±)-epicatechin in chocolate and cocoa. The method is the first such method developed by an industry group such as NCA for this purpose.

Abbreviations

HPLC:

High-Performance Liquid Chromatography

ISO:

International Organization for Standardization

mL:

Milliliter

°C:

Degrees Celsius

nm:

Nanometer

mm:

Millimeter

μm:

micrometer

PVDF:

Polyvinylidene Fluoride

PTFE:

Polytertrafluoroethylene

L:

Liter

mg:

Milligram

g:

Gram

Min:

Minutes

μg:

Microgram

%:

Percent

NIST:

National Institute of Standards and Technology

SRM:

Standard Reference Method

LOD:

Limit of Detection

LOQ:

Limit of Quantification

AOAC:

Association of Official Analytical Chemists

UV:

Ultraviolet

References

  1. Powis TG, Hurst WJ, Rodriguez MC, Ortiz CP, Blake M, Cheetham D, Coe MD, Hodgson JG: The Origins of Cacao Use in Mesoamerica. Mexicon. 2008, 30 (2): 35-38.

    Google Scholar 

  2. Crown PL, Hurst WJ: Evidence of Cacao Use in the Prehispanic American Southwest. Proc Natl Acad Sci USA. 2009, 106 (7): 2110-3. 10.1073/pnas.0812817106.

    Article  CAS  Google Scholar 

  3. Farouque H, Leung M, Hope S, Baldi M, Schechter C, Cameron J, Meredith I: Acute and Chronic Effects of Flavanol-Rich Cocoa on Vascular Function in Subjects With Coronary Artery Disease: a Randomized Double-Blind Placebo-Controlled Study. Clinical Science. 2006, 111: 71-80. 10.1042/CS20060048.

    Article  CAS  Google Scholar 

  4. Flammer AJ, Hermann F, Sudano I, Spieker L, Hermann M, Cooper KA, Serafini M, Luscher TF, Ruschitzka F, Noll G, Corti R: Dark Chocolate Improves Coronary Vasomotion and Reduces Platelet Reactivity. Circulation. 2007, 116 (21): 2376-2382. 10.1161/CIRCULATIONAHA.107.713867.

    Article  Google Scholar 

  5. Corti R, Flammer AJ, Hollenberg NK, Luscher TF: Cocoa and Cardiovascular Health. Circulation. 2009, 119 (10): 1433-1441. 10.1161/CIRCULATIONAHA.108.827022.

    Article  Google Scholar 

  6. Heiss C, Dejam A, Kleinbongard P, Schewe T, Sies H, Kelm M: Vascular Effects of Cocoa Rich in Flavan-3-Ols. Journal of American Medical Association. 2003, 290 (8): 1030-1031. 10.1001/jama.290.8.1030.

    Article  Google Scholar 

  7. Harborne JB: Phytochemical methods: a guide to modern techniques of plant analysis. 1998, Springer: New Delhi, India

    Google Scholar 

  8. Merken HM, Beecher GR: Liquid Chromatographic Method for the Separation and Quantification of Prominent Flavonoid Aglycones. Journal of Chromatography. 2000, 897 (1-2): 177-184. 10.1016/S0021-9673(00)00826-8.

    Article  CAS  Google Scholar 

  9. Moon JK, Shibamoto T: Antioxidant Assays for Plant and Food Components. J Agric Food Chem. 2009, 57 (5): 1655-1666. 10.1021/jf803537k.

    Article  CAS  Google Scholar 

  10. Beekwilder J, Jonker H, Meesters P, Hall RD, van der Meer IM, De Vos CHR: Antioxidants in Raspberry: on-Line Analysis Links Antioxidant Activity to a Diversity of Individual Metabolites. J Agric Food Chem. 2005, 53 (9): 3313-3320. 10.1021/jf047880b.

    Article  CAS  Google Scholar 

  11. Stewart AJ, Mullen W, Crozier A: On-Line High-Performance Liquid Chromatography Analysis of the Antioxidant Activity of Phenolic Compounds in Green and Black Tea. Mol Nutr Food Res. 2005, 49 (1): 52-60. 10.1002/mnfr.200400064.

    Article  CAS  Google Scholar 

  12. Gotti R, Fiori J, Mancini F, Cavrini V: Modified Micellar Electrokinetic Chromatography in the Analysis of Catechins and Xanthines in Chocolate. Electrophoresis. 2004, 25 (18-19): 3282-91.

    Article  CAS  Google Scholar 

  13. Hammerstone , Lazarus , Mitchell , Rucker , Schmitz : Identification of Procyanidins in Cocoa (Theobroma cacao) and Chocolate Using High Performance Liquid Chromatography/Mass Spectrometry. J Agric Food Chem. 1999, 47: 490-496. 10.1021/jf980760h.

    Article  CAS  Google Scholar 

  14. International Organization for Standardization: Determination of Substances Characterisic of Green and Black Tea -- Part 2: Content of Catechins in Green Tea -- Method Using High-Performance Liquid Chromatography. 2005, Geneva, Switzerland: ISO 14502-2

    Google Scholar 

  15. INA International: Determination of Total Catechin and Gallic Acid in Green Tea. [http://www.nsf.org/business/ina/greentea.asp?program=INA]

  16. Dean RB, Dixon WJ: Simplified Statistics for Small Numbers of Observations. Anal Chem. 1951, 23 (4): 636-638. 10.1021/ac60052a025.

    Article  CAS  Google Scholar 

  17. Swartz M, Krull I: Analytical Method Development and Validation. 1997, Marcell Dekker: New York, 21-35.

    Google Scholar 

  18. Serafini M, Bugianesi R, Maiani G, Valtuena S, De Santis S, Crozier A: Plasma antioxidants from chocolate. Nature. 2003, 424 (6952): 1013-10.1038/4241013a.

    Article  CAS  Google Scholar 

  19. The United States Pharmacopia: About the Columns Data. [http://www.usp.org/USPNF/columnsIntro.html]

  20. Alder L, Holland PT, Lantos J, Lee M, MacNeil JD, O'Rangers J, van Zoonen P, Ambrus A: Guidelines for Single-Laboratory Validation of Analytical Methods for Trace-Level Concentrations of Organic Chemicals. Principles and Practices of Method Validation. Edited by: Fajgelj A, Ambrus A. 2000, The Royal Society of Chemistry: Cambridge, UK, 179-248.

    Google Scholar 

  21. Wallace TC, Giusti MM: Extraction and Normal-Phase HPLC-Fluorescence-Electrospray MS Characterization and Quantification of Procyanidins in Cranberry Extracts. J Food Sci. 2010, 75 (8): C690-6. 10.1111/j.1750-3841.2010.01799.x.

    Article  CAS  Google Scholar 

  22. Unachukwu UJ, Ahmed S, Kavalier A, Lyles JT, Kennelly EJ: White and green teas (Camellia sinensis var. sinensis): variation in phenolic, methylxanthine, and antioxidant profiles. J Food Sci. 2010, 75 (6): C541-8. 10.1111/j.1750-3841.2010.01705.x.

    Article  CAS  Google Scholar 

  23. Sparzak B, Merino-Arevalo M, Vander Heyden Y, Krauze-Baranowska M, Majdan M, Fecka I, Głód D, Bączek T: HPLC analysis of polyphenols in the fruits of Rubus idaeus L. (Rosaceae). Nat Prod Res. 2010, 24 (19): 1811-22. 10.1080/14786411003754231.

    Article  CAS  Google Scholar 

  24. Furuuchi R, Yokoyama T, Watanabe Y, Hirayama M: Identification and quantification of short oligomeric proanthocyanidins and other polyphenols in boysenberry seeds and juice. J Agric Food Chem. 2011, 59 (8): 3738-46. 10.1021/jf104976n.

    Article  CAS  Google Scholar 

  25. Chira K, Lorrain B, Ky I, Teissedre PL: Tannin composition of cabernet-sauvignon and merlot grapes from the bordeaux area for different vintages (2006 to 2009) and comparison to tannin profile of five 2009 vintage mediterranean grapes varieties. Molecules. 2011, 16 (2): 1519-32. 10.3390/molecules16021519.

    Article  CAS  Google Scholar 

  26. Leitao C, Marchioni E, Bergaentzlé M, Zhao M, Didierjean L, Taidi B, Ennahar S: Effects of processing steps on the phenolic content and antioxidant activity of beer. J Agric Food Chem. 2011, 59 (4): 1249-55. 10.1021/jf104094c.

    Article  CAS  Google Scholar 

  27. Neves AC, Spranger MI, Zhao Y, Leandro MC, Sun B: Effect of addition of commercial grape seed tannins on phenolic composition, chromatic characteristics, and antioxidant activity of red wine. J Agric Food Chem. 2010, 58 (22): 11775-82. 10.1021/jf1038037.

    Article  CAS  Google Scholar 

Download references

Acknowledgements

The authors would like to thank the following company's laboratories for their participation in validating the method by testing commercial samples: The Hershey Company, Archer Daniels Midland Company, Kraft Europe, Kraft-Cadbury and Barry Callebaut. NCA is especially grateful to the following individuals for their commitment throughout this process: Jeff Hurst, Mark Payne, Lindo Groff, Mark Collison, Eva-Maria Berndt, Verena Jendreizik, Alison Branch and Olivier Nuytten.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Laura Shumow.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors' contributions

AB selected standard and sample preparation methods, chromatography parameters and suitable commercial samples. LS distributed samples to participating laboratories, collected and analyzed data and prepared the manuscript. All authors read and approved the final manuscript.

Authors’ original submitted files for images

Below are the links to the authors’ original submitted files for images.

Authors’ original file for figure 1

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License (https://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Shumow, L., Bodor, A. An industry consensus study on an HPLC fluorescence method for the determination of (±)-catechin and (±)-epicatechin in cocoa and chocolate products. Chemistry Central Journal 5, 39 (2011). https://doi.org/10.1186/1752-153X-5-39

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/1752-153X-5-39

Keywords